Algorithm for Turning Detection and Analysis Validated under Home-Like Conditions in Patients with Parkinson's Disease and Older Adults using a 6 Degree-of-Freedom Inertial Measurement Unit at the Lower Back

期刊

FRONTIERS IN NEUROLOGY
卷 8, 期 -, 页码 -

出版社

FRONTIERS MEDIA SA
DOI: 10.3389/fneur.2017.00135

关键词

six degrees of freedom; accelerometer; daily activities; gyroscope; older adults; Parkinson's disease; turning

资金

  1. EU [288557]
  2. Lundbeck

向作者/读者索取更多资源

Introduction: Aging and age-associated disorders such as Parkinson's disease (PD) are often associated with turning difficulties, which can lead to falls and fractures. Valid assessment of turning and turning deficits specifically in non-standardized environments may foster specific treatment and prevention of consequences. Methods: Relative orientation, obtained from 3D-accelerometer and 3D-gyroscope data of a sensor worn at the lower back, was used to develop an algorithm for turning detection and qualitative analysis in PD patients and controls in non-standardized environments. The algorithm was validated with a total of 2,304 turns >= 90 degrees extracted from an independent dataset of 20 PD patients during medication ON-and OFF-conditions and 13 older adults. Video observation by two independent clinical observers served as gold standard. Results: In PD patients under medication OFF, the algorithm detected turns with a sensitivity of 0.92, a specificity of 0.89, and an accuracy of 0.92. During medication ON, values were 0.92, 0.78, and 0.83. In older adults, the algorithm reached validation values of 0.94, 0.89, and 0.92. Turning magnitude (difference, 0.06 degrees; SEM, 0.14 degrees) and duration (difference, 0.004 s; SEM, 0.005 s) yielded high correlation values with gold standard. Overall accuracy for direction of turning was 0.995. Intra class correlation of the clinical observers was 0.92. Conclusion: This wearable sensor-and relative orientation-based algorithm yields very high agreement with clinical observation for the detection and evaluation of >= 90 degrees turns under non-standardized conditions in PD patients and older adults. It can be suggested for the assessment of turning in daily life.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据