期刊
SCIENTIA PHARMACEUTICA
卷 85, 期 1, 页码 -出版社
MDPI
DOI: 10.3390/scipharm85010006
关键词
amoxicillin; enrofloxacin; injectable suspension; HPLC; method development; validation
资金
- Advanced Veterinary Manufacturing Company (Palestine)
The combination of amoxicillin and enrofloxacin is a well-known mixture of veterinary drugs; it is used for the treatment of Gram-positive and Gram-negative bacteria. In the scientific literature, there is no high-performance liquid chromatography (HPLC)-UV method for the simultaneous determination of this combination. The objective of this work is to develop and validate an HPLC method for the determination of this combination. In this regard, a new, simple and efficient reversed-phase HPLC method for simultaneous qualitative and quantitative determination of amoxicillin and enrofloxacin, in an injectable preparation with a mixture of inactive excipients, has been developed and validated. The HPLC separation method was performed using a reversed-phase (RP)-C18e (250 mm x 4.0 mm, 5 mu m) column at room temperature, with a gradient mobile phase of acetonitrile and phosphate buffer containing methanol at pH 5.0, a flow rate of 0.8 mL/min and ultraviolet detection at 267 nm. This method was validated in accordance with the Food and Drug Administration (FDA) and the International Conference on Harmonisation (ICH) guidelines and showed excellent linearity, accuracy, precision, specificity, robustness, ruggedness, and system suitability results within the acceptance criteria. A stability-indicating study was also carried out and indicated that this method can also be used for purity and degradation evaluation of these formulations.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐