4.4 Article

Monitoring Pb in Aqueous Samples by Using Low Density Solvent on Air-Assisted Dispersive Liquid-Liquid Microextraction Coupled with UV-Vis Spectrophotometry

出版社

SPRINGER
DOI: 10.1007/s00128-016-2010-5

关键词

Air-assisted dispersive liquid-liquid microextraction; UV-Vis spectrophotometry; Chemometrics; Validation study; Lead; Aqueous samples

向作者/读者索取更多资源

In this study, AA-DLLME combined with UV-Vis spectrophotometry was developed for pre-concentration, microextraction and determination of lead in aqueous samples. Optimization of the independent variables was carried out according to chemometric methods in three steps. According to the screening and optimization study, 86 mu L of 1-undecanol (extracting solvent), 12 times syringe pumps, pH 2.0, 0.00% of salt and 0.1% DDTP (chelating agent) were chosen as the optimum independent variables for microextraction and determination of lead. Under the optimized conditions, R = 0.9994, and linearity range was 0.01-100 A mu g mL(-1). LOD and LOQ were 3.4 and 11.6 ng mL(-1), respectively. The method was applied for analysis of real water samples, such as tap, mineral, river and waste water.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据