A new method for rapid screening of unknown organic iodine (OI) in small-volume complex biological samples was developed using in-tube solid phase micro extraction (SPME) nanospray mass spectrometry (MS). The method proposed anew identification scheme for OI based on nanospray high-resolution mass spectrometry (HR-MS). The mass ranges of OI ions were confirmed using the t-MS2 scan mode first; then, the possible precursor ions of OI were selected and identified orderly in full MS/ddMS(2) and t-MS2 scan modes. Besides, in-tube SPME was used for. the pretreatment of small-volume biological samples, and it was the first time in-tube SPME combined with nanospray MS for OI identification. The whole analysis procedure took only 8 min and consumed 50 mu L per sample. Using the new method, six kinds of OI added to urine and an unknown OI C12H23O11I in human milk were successfully identified. Moreover, the proposed identification scheme is also suitable for other ambient mass spectrometry (AMS) to determine unknown compounds with characteristic fragment ions.