4.7 Article

A Urea Potentiometric Biosensor Based on a Thiophene Copolymer

期刊

BIOSENSORS-BASEL
卷 7, 期 1, 页码 -

出版社

MDPI
DOI: 10.3390/bios7010013

关键词

urea; urease; biosensors; potentiometry; polythiophene; conducting polymer

资金

  1. Kingston University

向作者/读者索取更多资源

A potentiometric enzyme biosensor is a convenient detector for quantification of urea concentrations in industrial processes, or for monitoring patients with diabetes, kidney damage or liver malfunction. In this work, poly(3-hexylthiophene-co-3-thiopheneacetic acid) (P(3HT-co-3TAA)) was chemically synthesized, characterized and spin-coated onto conductive indium tin oxide (ITO) glass electrodes. Urease (Urs) was covalently attached to the smooth surface of this copolymer via carbodiimide coupling. The electrochemical behavior and stability of the modified Urs/P(3HT-co-3TAA)/ITO glass electrode were investigated by cyclic voltammetry, and the bound enzyme activity was confirmed by spectrophotometry. Potentiometric response studies indicated that this electrode could determine the concentration of urea in aqueous solutions, with a quasi-Nernstian response up to about 5 mM. No attempt was made to optimize the response speed; full equilibration occurred after 10 min, but the half-time for response was typically <1 min.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据