Preparation of organic monolithic columns in polytetrafluoroethylene tubes for reversed-phase liquid chromatography

期刊

ANALYTICA CHIMICA ACTA
卷 960, 期 -, 页码 160-167

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.aca.2017.01.012

关键词

Polymer monolith; Polytetrafluoroethylene; Surface modification; Monolith attachment; Microbore column; Reversed-phase liquid chromatography

资金

  1. MINECO of Spain and Fondo Europeo de Desarrollo Regional, FEDER [CTQ2014-52765-R]
  2. Conselleria d'Educacio, Investigacio, Cultura i Esport, Generalitat Valenciana, Spain [PROMETEO/2016/145]

向作者/读者索取更多资源

In this work, a method for the preparation and anchoring of polymeric monoliths in a polytetrafluoroethylene (PTFE) tubing as a column housing for microbore HPLC is described. In order to assure a covalent attachment of the monolith to the inner wall of the PTFE tube, a two-step procedure was developed. Two surface etching reagents, a commercial sodium naphthalene solution (Fluoroetch (R)), or mixtures of H2O2 and H2SO4, were tried and compared. Then, the obtained hydroxyl groups on the PTFE surface were modified by methacryloylation. Attenuated total reflectance Fourier-transform infrared (ATR-FTIR) spectroscopy and scanning electron microscopy (SEM) confirmed the successful modification of the tubing wall and the stable anchorage of monolith to the wall, respectively. Special emphasis was also put on the reduction of the unwanted effects of shrinking of monolith during polymerization, by using an external proper mold and by selecting the adequate monomers in order to increase the flexibility of the polymer. Poly(glycidyl methacrylate-co-divinylbenzene) monoliths were in situ synthesized by thermal polymerization within the confines of surface-vinylized PTFE tubes. The modified PTFE tubing tightly held the monolith, and the monolithic column exhibited good pressure resistance up to 20 MPa. The column performance was also evaluated via the isocratic separation of a series of alkylbenzenes in the reversed-phase mode. The optimized monolithic columns gave plate heights ranged between 70 and 80 Am. The resulting monoliths were also satisfactorily applied to the separation of proteins. (C) 2017 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据