4.8 Article

Temperature and pH sensors based on graphenic materials

期刊

BIOSENSORS & BIOELECTRONICS
卷 91, 期 -, 页码 870-877

出版社

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2017.01.062

关键词

pH; Temperature; Graphene; Flexible sensor

资金

  1. EU [FP7 ICT-317894]
  2. Ente Cassa di Risparmio di Firenze

向作者/读者索取更多资源

Point-of-care applications and patients' real-time monitoring outside a clinical setting would require disposable and durable sensors to provide better therapies and quality of life for patients. This paper describes the fabrication and performances of a temperature and a pH sensor on a biocompatible and wearable board for healthcare applications. The temperature sensor was based on a reduced graphene oxide (rGO) layer that changed its electrical resistivity with the temperature. When tested in a human serum sample between 25 and 43 C-omicron, the sensor had a sensitivity of 110 +/- 10 Omega/ C-omicron and an error of 0.4 +/- 0.1 C-omicron compared with the reference value set in a thermostatic bath. The pH sensor, based on a graphene oxide (GO) sensitive layer, had a sensitivity of 40 4 mV/pH in the pH range between 4 and 10. Five sensor prototypes were tested in a human serum sample over one week and the maximum deviation of the average response from reference values obtained by a glass electrode was 0.2 pH units. For biological applications, the temperature and pH sensors were successfully tested for in vitro cytotoxicity with human fibroblast cells (MRC-5) over 24 h.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据