4.5 Article

Rapid mixing with high-throughput in a semi-active semi-passive micromixer

期刊

ELECTROPHORESIS
卷 38, 期 9-10, 页码 1310-1317

出版社

WILEY
DOI: 10.1002/elps.201600393

关键词

Active mixer; Electrothermal; Microgrooved channel; Passive mixer

向作者/读者索取更多资源

In this paper, we investigate a novel alternating current electrothermal ( ACET) micromixer driven by a high efficiency ACET micropump. The micromixer consists of thin film asymmetric pairs of electrodes on the microgrooved channel floor and array of electrode pairs fabricated on the top wall. By connecting electrodes with AC voltage, ACET forces are induced. Asymmetric microgrooved electrodes force the fluids along the channel, while lateral vortex pairs are generated by symmetric electrode pairs located on the top wall. Waviness of the floor increases contact area between two confluent streams within a narrow confinement. An active mixer operates as a semi active semi passive mixer. Effects of various parameters are investigated in details in order to arrive at an optimal configuration that provides for efficient mixing as well as appreciable transport. It is found that using a specific design, uniform and homogeneousmixing quality with mixing efficiency of 97.25% and flow rate of 1.794 mu m(2)/min per unit width of the channel can be achieved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据