CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS
卷 164, 期 -, 页码 52-55出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.chemolab.2017.03.012
关键词
standard addition; calibration; bias; zero gas; matrix
类别
资金
- National Research Foundation of Korea (NRF) - Ministry of Science, ICT Future Planning [2016R1E1A1A01940995]
- UK National Measurement System
A fixed volume case of sequential standard addition calibration (S-SAC) for the value assignment of impurities in zero gas is described. A mathematical description of this technique has been derived and has been shown to exhibit a similar systematic bias during the extrapolation process to that seen for other S-SAC cases. The use of S-SAC in the gas phase has demonstrated that variation in the sample volume for S-SAC in the liquid phase is not the generalised cause of the bias experienced during extrapolation. Instead it is the variation in the volume of the original sample with respect to the overall volume of the mixture following the addition of standard. In addition, the requirement to perform a bias correction on the extrapolated values has been discussed and best practice solutions for correction proposed.