4.7 Article

Multicenter evaluation of an enzymatic method for glycated albumin

期刊

CLINICA CHIMICA ACTA
卷 469, 期 -, 页码 81-86

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.cca.2017.03.028

关键词

Diabetes; Glycated albumin; Multicenter evaluation

向作者/读者索取更多资源

Background: The use of glycated albumin (GA) has been proposed as an additional glycemic control marker particularly useful in intermediate-term monitoring and in situation when HbA(1c), test is not reliable. Methods: We have performed the first multicenter evaluation of the analytical performance of the enzymatic method quantlLab Glycated Albumin assay implemented on the most widely used clinical chemistry analyzers (i.e. Abbott Architect C8000, Beckman Coulter AU 480 and 680, Roche Cobas C6000, Siemens ADVIA 2400 and 2400 XPT). Results: The repeatability of the GA measurement (expressed as CV, %) implemented in the participating centers ranged between 0.9% and 1.2%. The within-laboratory CVs ranged between 1.2% and 1.6%. A good alignment between laboratories was found, with correlation coefficients from 0.996 to 0.998. Linearity was confirmed in the range from 7.6 to 84.7%. Conclusion: The new enzymatic method for glycated albumin evaluated by our investigation is suitable for clinical use.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据