Validation of a hand-held point of care device for lactate in adult and pediatric patients using traditional and locally-smoothed median and maximum absolute difference curves

期刊

CLINICA CHIMICA ACTA
卷 468, 期 -, 页码 145-149

出版社

ELSEVIER
DOI: 10.1016/j.cca.2017.02.015

关键词

Lactate; Point-of-care; Trauma; Sepsis; Pre-hospital; Validation; Locally-smoothed median absolute difference curve; Locally-smoothed maximum absolute difference curve

资金

  1. Nova Biomedical

向作者/读者索取更多资源

Background: Lactate is commonly used in septic patients and is a viable biomarker for trauma patients. Its pre-hospital use could assist triaging and managing patients with these conditions. Methods: We evaluated the analytical performance of the point-of-care (POC) StatStrip Xpress Lactate Meter (Nova Biomedical) and compared it to the ABL 800 (Radiometer). We measured lactate in 250 adult and 250 pediatric whole blood samples in 2 laboratories. The performance of the POC meter was assessed by traditional linear regression and Bland-Altman plots, and locally-smoothed (LS) median absolute difference and maximum absolute difference (MAD and MaxAD) curves. Results: The StatStrip was linear with acceptable reproducibility at clinically relevant concentrations. Correlation with the ABL800 showed a negative bias for both populations with slope, bias +/- SD (% bias) of 0.78, -0.4 +/- 0.7 (-14.5%) in children and 0.80-03 +/- 0.6 (- 13.3%) in adults. The proportional bias appeared more significant at concentrations >4 mmol/l (36.0 mg/dl). The StatStrip misclassified 7.6 and 8.8% pediatric and adult samples, respectively, to lower risk categories defined using guidelines driven cut-offs. The LS MAD curves identified one breakout, concentration where the LS MAD exceeds the total allowable error limit of 03 mmol/l (2.7 mg/dl), at lactate concentrations of 3.8 and 3.2 mmol/l (34.2 and 28.8 mg/dl) in the pediatric and adult curves, respectively. Breakthroughs, points at which the LS MaxAD curve exceeds the 95th percentile of MaxADs, occur at concentrations above 7.5 mmol/l (67.6 mg/dl) for both populations where the performance of the POC meter became erratic. We concluded that if serial lactate measurements are performed, the same method should be used for baseline and follow up measurements. The LS MAD and LS MaxAD curves allowed visual and quantitative mapping of the performance of the lactate POC meter over the range of concentrations measured. Conclusions: This approach seems useful for the identification of points at which the performance of a POC meter differs significantly from a comparison method and thresholds of poor analytical performance. (C) 2017 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据