Accuracy of a Fourth-Generation Subcutaneous Continuous Glucose Sensor

期刊

DIABETES TECHNOLOGY & THERAPEUTICS
卷 19, 期 8, 页码 446-456

出版社

MARY ANN LIEBERT, INC
DOI: 10.1089/dia.2017.0087

关键词

Glucose sensor; Sensor-integrated pump; Sensor accuracy; Mobile device; MARD

资金

  1. Medtronic

向作者/读者索取更多资源

Background: This study evaluated the accuracy and performance of a fourth-generation subcutaneous glucose sensor (Guardian (TM) Sensor 3) in the abdomen and arm. Methods: Eighty-eight subjects (14-75 years of age, mean +/- standard deviation [SD] of 42.0 +/- 19.1 years) with type 1 or type 2 diabetes participated in the study. Subjects wore two sensors in the abdomen that were paired with either a MiniMed (TM) 640G insulin pump, or an iPhone (R) or iPod (R) touch (R) running a glucose monitoring mobile application (Guardian Connect system) and a third sensor in the arm, which was connected to a glucose sensor recorder (GSR). Subjects were also asked to undergo in-clinic visits of 12-14 h on study days 1, 3, and 7 for frequent blood glucose sample testing using a Yellow Springs Instrument (YSI) reference. Results: The overall mean absolute relative difference (MARD-SD) between abdomen sensor glucose (SG) and YSI reference values was 9.6% +/- 9.0% and 9.4% +/- 9.8% for the MiniMed 640G insulin pump and Guardian Connect system, respectively; and 8.7% +/- 8.0% between arm SG and YSI reference values. The percentage of SG values within 20% agreement of the YSI reference value (for YSI > 80 mg/dL) was 90.7% with the MiniMed 640G insulin pump, 91.8% with the Guardian Connect system, and 93.1% for GSR-connected arm sensors. Mean functional sensor life, when calibrating 3-4 times/day, was 145.9 +/- 39.3 h for sensors paired with the MiniMed 640G insulin pump, 146.1 +/- 41.6 h for sensors paired with the Guardian Connect system, and 147.6 +/- 40.4 h for sensors connected to the GSR. Responses to survey questions regarding sensor comfort and ease of use were favorable. Conclusions: The Guardian Sensor 3 glucose sensor, whether located in abdomen or the arm, provided accurate glucose readings when compared with the YSI reference and demonstrated functional life commensurate with the intended 7-day use.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据