4.1 Article

Slice-accelerated gradient-echo echo planar imaging dynamic susceptibility contrast-enhanced MRI with blipped CAIPI: effect of increasing temporal resolution

期刊

JAPANESE JOURNAL OF RADIOLOGY
卷 36, 期 1, 页码 40-50

出版社

SPRINGER
DOI: 10.1007/s11604-017-0695-y

关键词

Blipped CAIPI; DSC; Perfusion; SMS; Temporal resolution

资金

  1. ImPACT Program of the Council for Science, Technology and Innovation (Cabinet Office, Government of Japan)

向作者/读者索取更多资源

To assess the influence of high temporal resolution on the perfusion measurements and image quality of perfusion maps, by applying simultaneous-multi-slice acquisition (SMS) dynamic susceptibility contrast-enhanced (DSC) magnetic resonance imaging (MRI). DSC-MRI data using SMS gradient-echo echo planar imaging sequences in 10 subjects with no intracranial abnormalities were retrospectively analyzed. Three additional data sets with temporal resolution of 1.0, 1.5, and 2.0 s were created from the raw data sets of 0.5 s. Cerebral blood flow (CBF), cerebral blood volume, mean transit time (MTT), time to peak (TTP), and time to maximum tissue residue function (T (max)) measurements were performed, as was visual perfusion map analysis. The perfusion parameter for temporal resolution of 0.5 s (reference) was compared with each synthesized perfusion parameter. CBF, MTT, and TTP values at temporal resolutions of 1.5 and 2.0 s differed significantly from the reference. The image quality of MTT, TTP, and T (max) maps deteriorated with decreasing temporal resolution. The temporal resolution of DSC-MRI influences perfusion parameters and SMS DSC-MRI provides better image quality for MTT, TTP, and T (max) maps.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据