A Shape Memory Alloy-Based Compression Therapy Prototype Tested With Individuals in Seated Position

出版社

ASME
DOI: 10.1115/1.4037441

关键词

compression therapy; shape memory alloy; active compression brace; smart materials; orthostatic hypotension

资金

  1. Natural Sciences and Engineering Research Council of Canada (NSERC)
  2. Canadian Institutes of Health Research (CIHR)
  3. Canada Research Chairs (CRC)

向作者/读者索取更多资源

Orthostatic intolerance in patients can occur secondary to concomitant venous pooling and enhanced capillary filtration when standing upright, and is one of the principle causes of syncope or fainting. Compression therapy is commonly recommended for the management of syncope based on the assumption that it increases venous return. Technologies currently used include compression stockings, whose efficacy has, however, been challenged, and intermittent pneumatic pressure devices, which highly restrict the patients' mobility. This paper therefore investigates a novel active compression brace (ACB), which could potentially provide intermittent pressure while not restricting movements. The ACB, actuated by shape memory alloy (SMA) wires, in this work was tested with twelve healthy individuals in a seated position. The experimental observation showed that the ACB can apply a constant initial pressure to the leg similar to commercial compression stockings and also produce intermittent pressure exceeding 30 mmHg. A comparison between analytical and experimental results showed a maximum of 2.08 mmHg absolute averaged difference among all the participants. A correlation analysis showed that the normalized root-mean-square deviation (NRMSD) between the experimental and analytical results had a significant negative correlation with the estimated total calf circumference minus the calf fat cross-sectional area (CSA). A calibration formula, accounting for fat and circumference of the leg, was introduced to account for these two parameters. The comfort of the ACB was also compared to two other available compression devices using questionnaires. No participants reported discomfort in terms of pressure, skin irritation, or heat generated by the ACB.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据