3.8 Proceedings Paper

Measurement system analysis of viscometers used for drilling mud characterization

出版社

IOP PUBLISHING LTD
DOI: 10.1088/1757-899X/222/1/012003

关键词

-

资金

  1. Ministry of Education Malaysia through Faculty of Chemical Engineering, Universiti Teknologi MARA, Selangor [RAGS/1/2014/TK04/UITM//8]

向作者/读者索取更多资源

Viscometers in the Faculty of Chemical Engineering, University Teknologi MARA, are subject to heavy utilization from the members of the faculty. Due to doubts surrounding their result integrity and maintenance management, Measurement System Analysis was executed. 5 samples of drilling muds with varied barite content from 5 - 25 weight% were prepared and their rheological properties determined in 3 trials by 3 operators using the viscometers. Gage Linearity and Bias Study were performed using Minitab software and the result shows high biases in the range of 19.2% to 38.7%, with non-linear trend along the span of measurements. Gage Repeatability & Reproducibility (Nested) analysis later produces Percent Repeatability & Reproducibility more than 7.7% and Percent Tolerance above 30%. Lastly, good and marginal Distinct Categories output are seen among the results. Despite acceptable performance of the measurement system in Distinct Categories, the poor results in accuracy, linearity, and Percent Repeatability & Reproducibility render the gage generally not capable. Improvement to the measurement system is imminent.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据