BIOMEDICAL CHROMATOGRAPHY
卷 31, 期 11, 页码 -出版社
WILEY
DOI: 10.1002/bmc.4000
关键词
mouse plasma; PCB 180; PCB 77; pharmacokinetics; UPLC-dad
A simple, sensitive and reproducible ultra-performance liquid chromatography (UPLC) method has been developed and validated for simultaneous estimation of polychlorinated biphenyl (PCB) 77 and PCB 180 in mouse plasma. The sample preparation was performed by simple liquid-liquid extraction technique. The analytes were chromatographed on a Waters Acquity H class UPLC system using isocratic mobile phase conditions at a flow rate of 0.3 mL/min and Acquity UPLC BEH shield RP18 column maintained at 35 degrees C. Quantification was performed on a photodiode array detector set at 215nm and PCB 101 was used as internal standard (IS). PCB 77, PCB 180, and IS retention times were 2.6, 4.7 and 2.8min, respectively, and the total run time was 6min. The method was validated for specificity, selectivity, recovery, linearity, accuracy, precision and sample stability. The calibration curve was linear over the concentration range 10-3000 ng/mL for PCB 77 and PCB 180. Intra- and inter-day precisions for PCBs 77 and 180 were found to be good with CV <4.64%, and the accuracy ranged from 98.90 to 102.33% in mouse plasma. The validated UPLC method was successfully applied to the pharmacokinetic study of PCBs 77 and 180 in mouse plasma.