BACKGROUND: Bacterial contamination of blood components remains a major cause of sepsis in transfusion medicine. Between 2006 and 2010 in the 5 years before the introduction of bacterial screening of platelet (PLT) components by National Health Service Blood and Transplant (NHSBT), seven cases of PLT component-associated transmission of bacterial infection were recorded for 10 patients, three of which were fatal. STUDY DESIGN AND METHODS: Sampling of individual PLT components was undertaken at 36 to 48 hours after donation and tested in the BacT/ALERT system with 8 mL inoculated into each of aerobic and anaerobic culture bottles. Bottles were incubated until the end of the 7-day shelf life and initial reactive bottles were examined for contamination. Bacterial screened time-expired PLTs were tested as in the screen method. RESULTS: From February 2011 to September 2015, a total of 1,239,029 PLT components were screened. Initial-reactive, confirmed-positive, and false-positive rates were 0.37, 0.03, and 0.19%, respectively. False-negative cultures, all with Staphylococcus aureus, occurred on four occasions; three were visually detected before transfusion and one confirmed transmission resulted in patient morbidity. The NHSBT screening protocol effectively reduced the number of clinically adverse transfusion transmissions by 90% in this reporting period, compared to a similar time period before implementation. Delayed testing of 4515 time-expired PLT units after screening revealed no positives. CONCLUSION: The implementation of bacterial screening of PLT components with the NHSBT BacT/ALERT protocol was an effective risk reduction measure and increased the safety of the blood supply.