Sampling Error Profile Analysis for calibration transfer in multivariate calibration

期刊

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.chemolab.2017.10.016

关键词

Sampling Error Profile Analysis (SEPA); Calibration transfer; PDS; Near-infrared

资金

  1. Technology Center, Shanghai Tobacco Group Corporation, Key Laboratory of Cigarette Smoke Research of CNTC [K2015-1-031p]

向作者/读者索取更多资源

A new strategy named Sampling Error Profile Analysis (SEPA) is proposed in the optimization for some parameters in piecewise direct standardization (PDS), such as the number of principal components and window size, and the evaluation for the calibration transfer. Partial least squares (PLS) with mean-centering is used in PDS for calibration transfer. Random re-sampling is carried out in SEPA to obtain a series of subsets and build same number sub-models that produce corresponding number root mean square errors (RMSE), of which the mean value and standard deviation are calculated. To take both accuracy and stability into account, the sum of the mean value and standard deviation are used for parameter optimization and model evaluation. The performance of the proposed strategy has been tested on two data sets: a ternary mixture dataset and a corn dataset. Compared with PDS, SEPA-PDS obtained lower prediction errors, indicating that the transfer model would be more robust and effective when using the parameters optimized by SEPA. Compared with other two commonly used calibration transfer methods of slope and bias correction (SBC) and spectral space transformation (SST), SEPA-PDS acquired more satisfactory results.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据