Mega Acer Kit® is more effective for warming the intravenous fluid than Ranger™ and ThermoSens® at 440 ml/h of infusion rate: an experimental performance study

期刊

KOREAN JOURNAL OF ANESTHESIOLOGY
卷 70, 期 4, 页码 456-461

出版社

KOREAN SOC ANESTHESIOLOGISTS
DOI: 10.4097/kjae.2017.70.4.456

关键词

Fluid temperature; Intravenous fluid warmer; Low flow rate; Mega Acer Kit; Ranger; ThermoSens

资金

  1. Clinical Medicine Research Institute at Chosun University Hospital

向作者/读者索取更多资源

Background: Few studies have investigated the effectiveness of intravenous fluid warmers at low and moderate flow rates below 1,000 ml/h. In this study, we compared the effectiveness of three different fluid warmers at a low flow rate (440 ml/h). Methods: We experimentally investigated the fluid warming performances of Mega Acer Kit (R) (Group M, n = 10), Ranger (TM) (Group R, n = 10), and ThermoSens (R) (Group T, n = 10) at 440 ml/h for 60 min. All devices were set at a warming temperature of 41 degrees C with preheating for 10 min. Intravenous fluids were then delivered through them. The fluid temperature (primary endpoint) was measured at 76 cm from the device after infusion for 60 min. The expected decrease in mean body temperature (secondary endpoint) after 5 h infusion for a 70 kg patient (Delta MBT5) was also calculated. Results: The fluid temperature (mean [95% CI]) at 76 cm from the device, 60 minutes after the infusion was higher in group M (36.01[35.73-36.29]degrees C), compared to groups T (29.81 [29.38-30.24]degrees C) and R (29.12 [28.52-29.72]degrees C) (P < 0.001). The Delta MBT5 (mean [95% CI]) was significantly smaller in group M (-0.04 [-0.04 to -0.03]degrees C) than that in groups T (-0.27 [-0.28 to -0.29]degrees C; P < 0.001) and R (-0.30 [-0.32 to -0.27]degrees C; P < 0.001). However, none of the fluid warmers provided a constant normothermic temperature above 36.5 degrees C. Conclusions: Mega Acer Kit (R) was more effective in warming the intravenous fluid with the smallest expected change in the mean body temperature, compared to Ranger (TM) and ThermoSens (R), at a flow rate of 440 ml/h.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据