3.9 Article

Raman Spectroscopy System for Non-invasive Blood Glucose Detection

期刊

ACTA PHOTONICA SINICA
卷 46, 期 8, 页码 -

出版社

SCIENCE PRESS
DOI: 10.3788/gzxb20174608.0812005

关键词

Raman spectroscopy; Spectrometers; Nondestructive testing; Blood or tissue constituent monitoring; Blood glucose

类别

资金

  1. National Natural Science Foundation of China [61308086]
  2. Natural Science Basic Research Plan in Shaanxi Province of China [2016JQ6016]

向作者/读者索取更多资源

A miniature wearable Raman spectroscopy system used to achieve noninvasive detection of human blood glucose level was developed. A thallium-doped grin lens was employed as the collection lens and a specially designed wearable fiber optic probe was employed to help the stable and convenient collection of Raman spectrum. The glucose solution, 11 rats and 10 healthy human were studied as subjects. In addition, a method of quantitative analysis of Raman spectrum was proposed which using the peak area as the main reference factor while the peak intensity as the auxiliary reference factor to calculate the target concentration. A non-linearized multivariate dominant factor-based partial least squares model was built for different samples to predict glucose level. The results show that the accuracy are 98.1%, 89.3% and 84.4% for glucose solution, rats and human subjects. The system has the advantages of more compact structure, lower cost, better testing stability and convenient for human body to wear, and is feasible and repeatable to achieve the noninvasive detection of human blood glucose accurately.