Tetrazolium salt monomers for gel dosimetry II: Dosimetric characterization of the ClearView™ 3D dosimeter

出版社

IOP PUBLISHING LTD
DOI: 10.1088/1742-6596/847/1/012049

关键词

-

向作者/读者索取更多资源

ClearView is a colourless radiochromic 3D dosimeter that is chemically and mechanically stable before and after irradiation. Upon irradiation, colour is generated from the radiochemical reduction of a tetrazolium salt suspended within a gellan gum gel. Here we present the dosimetric characterization of small and large volume samples (4 mL and 1 L, respectively) from three different batches of ClearView. The dose sensitivity was linear up to 80 Gy and constant within a batch but varied among batches between 3.4x10(-3) and 4.1x10(-3) Gy(-1)cm(-1). Once generated, the radiation signal within the large samples remained unchanged for at least 58 days. Both electrons and photons produced the same signal at different energy levels; however, at dose rates below 400 cGy/min, a drop in sensitivity was observed. Overall, ClearView is best suited for relative dosimetry up to 80 Gy at high dose rates.