Development of novel amperometric urea biosensor based on Fc-PAMAM and MWCNT bio-nanocomposite film

期刊

SENSORS AND ACTUATORS B-CHEMICAL
卷 246, 期 -, 页码 920-926

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2017.02.122

关键词

Ferrocence; PAMAM dendrimer; Urea; Urease; Bionanocomposite; Amperometric biosensor; MWCNT

向作者/读者索取更多资源

In this work, fabrication of new electrochemical urea biosensor based on ferrocene-poly(amidoamine) (Fc-PAMAM) dendrimers incorporated with multi walled carbon nanotubes (MWCNTs) is reported. Bio-sensing electrode was characterized using cyclic voltammogram where optimization studies were performed for reusability, storage, interference behaviors together with optimum temperature and pH parameters. Electrode showed excellent performance in urea detection with wide linear range of 0.2-1.8 mM, detection limit of 0.05 mM, fast response of 3 s, and sensitivity of 1.085 mu A/cm(2)/mM. Final assessment of electrochemical biosensor was real sample application which was performed in healthy human blood sample where excellent performance in selectivity and sensitivity of the bio-electrode was observed. Fabricated bio-sensing electrode has great potential in urea measurement applications. (C) 2017 Published by Elsevier B.V.