4.7 Article

Conditional siphon priming for multi-step assays on centrifugal microfluidic platforms

期刊

SENSORS AND ACTUATORS B-CHEMICAL
卷 242, 期 -, 页码 710-717

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2016.11.063

关键词

Centrifugal microfluidics; Conditional siphon priming; Ammonium analysis

资金

  1. National Natural Science Foundation of China [31500691, 81471749, 81327005]
  2. Beijing Municipal Science & Technology Commission [Z161100000116031]
  3. Beijing Lab Foundation

向作者/读者索取更多资源

Centrifugal microfluidics has proven to be successful in biomedical diagnostics, biological analysis and environmental monitoring. However, the automation of multi-step sample processing, reaction and detection remains a great challenge. In this study, a conditional siphon priming technique is introduced for multi-step liquid addition or selective routing. Since the siphon channel is locally modified or venting is blocked by liquid in another chamber, siphon priming can be triggered by liquid addition or venting at low frequency. Using this technique, sequential release,of liquids and selective routing in multiple manners were successfully achieved. As a proof of concept, a centrifugal microfluidic platform was designed for on-site ammonium analysis in water samples. The linear range of ammonium concentrations is extended by integration of a dilution process. This novel valving technique provides new solutions for integration of complex liquid handling processes on centrifugal platforms. (C) 2016 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。