4.6 Article

Mid-Infrared Trace Gas Sensor Technology Based on Intracavity Quartz-Enhanced Photoacoustic Spectroscopy

期刊

SENSORS
卷 17, 期 3, 页码 -

出版社

MDPI
DOI: 10.3390/s17030513

关键词

laser absorption spectroscopy; bow-tie cavity; I-QEPAS; intracavity quartz-enhanced photoacoustic spectroscopy; mid-infrared trace gas sensors

资金

  1. Polish National Center for Research and Development [179616]
  2. National Science Foundation (NSF) ERC MIRTHE
  3. Robert Welch Foundation [C-0586]

向作者/读者索取更多资源

The application of compact inexpensive trace gas sensor technology to a mid-infrared nitric oxide (NO) detectoion using intracavity quartz-enhanced photoacoustic spectroscopy (I-QEPAS) is reported. A minimum detection limit of 4.8 ppbv within a 30 ms integration time was demonstrated by using a room-temperature, continuous-wave, distributed-feedback quantum cascade laser (QCL) emitting at 5.263 mu m (1900.08 cm(-1)) and a new compact design of a high-finesse bow-tie optical cavity with an integrated resonant quartz tuning fork (QTF). The optimum configuration of the bow-tie cavity was simulated using custom software. Measurements were performed with a wavelength modulation scheme (WM) using a 2f detection procedure.