4.5 Article

Low noise current preamplifier for qPlus sensor deflection signal detection in atomic force microscopy at room and low temperatures

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 88, 期 7, 页码 -

出版社

AMER INST PHYSICS
DOI: 10.1063/1.4993737

关键词

-

资金

  1. Deutsche Forschungsgemeinschaft [Sonderforschungsbereich 689]

向作者/读者索取更多资源

The resolution of frequency modulation atomic force microscopy is limited by instrumental noise. When using a qPlus sensor, the deflection detector noise is the dominant noise contribution. It can be reduced by improving the preamplifier used to amplify the sensor deflection signal. We present a simple single-stage differential preamplifier which outperforms previous designs known to us by at least a factor of two in the deflection noise density. We show specific versions of this preamplifier to use in ambient conditions, in ultra-high vacuum at room temperature, and at 4.2 K. Furthermore, we compare the thermal peak analysis and the frequency shift noise density method as a means to determine the deflection noise density. We note that this preamplifier can also be used for any current-generating sensors such as other piezoelectric sensors and photodiodes, but, in this paper, we restrict our analysis to qPlus sensors. Published by AIP Publishing.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据