4.7 Article

Simultaneous determination of acetaminophen and oxycodone in human plasma by LC-MS/MS and its application to a pharmacokinetic study

期刊

JOURNAL OF PHARMACEUTICAL ANALYSIS
卷 8, 期 3, 页码 160-167

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jpha.2018.01.006

关键词

Acetaminophen; Oxycodone; LC-MS/MS; Human plasma; Pharmacokinetics

向作者/读者索取更多资源

A simple and rapid liquid chromatography tandem mass spectrometry (LC-MS/MS) method was developed and validated for simultaneous determination of acetaminophen and oxycodone in human plasma. Acetaminophen-d4 and oxycodone-d3 were used as internal standards. The challenge encountered in the method development that the high plasma concentration level of acetaminophen made the MS response saturated while the desired lower limit of quantification (LLOQ) for oxycodone was hard to reach was well solved. The analytes were extracted by protein precipitation using acetonitrile. The matrix effect of the analytes was avoided by chromatographic separation using a hydrophilic C-18 column coupled with gradient elution. Multiple reaction monitoring in positive ion mode was performed on tandem mass spectrometer employing electrospray ion source. The calibration curves were linear over the concentration ranges of 40.0-8000 ng/mL and 0.200-40.0 ng/mL for acetaminophen and oxycodone, respectively. This method, which could contribute to high throughput analysis and better clinical drug monitoring, was successfully applied to a pharmacokinetic study in healthy Chinese volunteers. (C) 2018 Xi'an Jiaotong University. Production and hosting by Elsevier B.V.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据