3.8 Article

Novel Mechanomyogram/electromyogram Hybrid Transducer Measurements Reflect Muscle Strength during Dynamic Exercise - Pedaling of Recumbent Bicycle -

期刊

ADVANCED BIOMEDICAL ENGINEERING
卷 7, 期 -, 页码 47-54

出版社

JAPANESE SOC MEDICAL & BIOLOGICAL ENGINEERING
DOI: 10.14326/abe.7.47

关键词

mechanomyogram (MMG); electromyogram (EMG); displacement-MMG; recumbent bicycle

资金

  1. Okayama Prefecture
  2. Japan Society for the Promotion of Science [17K01360]
  3. Grants-in-Aid for Scientific Research [17K01360] Funding Source: KAKEN

向作者/读者索取更多资源

Simultaneous measurements of mechanomyogram (MMG) and electromyogram (EMG) may be useful for accurate evaluation of skeletal muscle contraction. However, unlike the EMG, the MMG is rarely used in clinical tests. As the target muscle has to be fixed during conventional MMG measurements, it is not possible to measure MMG during dynamic exercises such as sports and rehabilitation. To solve these problems, the authors developed an MMG (displacement-MMG)/EMG hybrid transducer system that allows simultaneous MMG and EMG measurements. Furthermore, we also developed an analysis method that is able to evaluate muscle contraction using the power spectra of each signal. The measurement system and the analysis method were applied to recumbent bicycle pedaling, during which work rate was increased incrementally. The results showed that this transducer system provided MMG/EMG measurements stably during dynamic exercise. When the work rate of the bicycle pedaling increased; that is, when the dynamic muscle strength increased, the sum of the power spectra of the MMG/EMG also increased. The MMG/EMG hybrid transducer system and the analysis method were useful for evaluating muscle strength during dynamic exercise.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据