Regional deep hyperthermia: impact of observer variability in CT-based manual tissue segmentation on simulated temperature distribution

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 62, 期 11, 页码 4479-4495

出版社

IOP PUBLISHING LTD
DOI: 10.1088/1361-6560/aa685b

关键词

deep regional hyperthermia; temperature distribution; manual segmentation; observer variability; hyperthermia treatment planning

资金

  1. German Federal Ministry for Economic Affairs and Energy [ZIM KF2769328KM4]

向作者/读者索取更多资源

The aim of this study was to systematically investigate the influence of the inter-and intra-observer segmentation variation of tumors and organs at risk on the simulated temperature coverage of the target. CT scans of six patients with tumors in the pelvic region acquired for radiotherapy treatment planning were used for hyperthermia treatment planning. To study the effect of interobserver variation, three observers manually segmented in the CT images of each patient the following structures: fat, muscle, bone and the bladder. The gross tumor volumes (GTV) were contoured by three radiation oncology residents and used as the hyperthermia target volumes. For intra-observer variation, one of the observers of each group contoured the structures of each patient three times with a time span of one week between the segmentations. Moreover, the impact of segmentation variations in organs at risk (OARs) between the three inter-observers was investigated on simulated temperature distributions using only one GTV. The spatial overlap between individual segmentations was assessed by the Dice similarity coefficient (DSC) and the mean surface distance (MSD). Additionally, the temperatures T90/T10 delivered to 90%/10% of the GTV, respectively, were assessed for each observer combination. The results of the segmentation similarity evaluation showed that the DSC of the inter-observer variation of fat, muscle, the bladder, bone and the target was 0.68 +/- 0.12, 0.88 +/- 0.05, 0.73 +/- 0.14, 0.91 +/- 0.04 and 0.64 +/- 0.11,respectively. Similar results were found for the intra-observer variation. The MSD results were similar to the DSCs for both observer variations. A statistically significant difference (p < 0.05) was found for T90 and T10 in the predicted target temperature due to the observer variability. The conclusion is that intra-and inter-observer variations have a significant impact on the temperature coverage of the target. Furthermore, OARs, such as bone and the bladder, may essentially influence the homogeneity of the simulated target temperature distribution.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据