4.4 Review

Dose calculation algorithm accuracy for small fields in non-homogeneous media: The lung SBRT case

期刊

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.ejmp.2016.11.104

关键词

Lung SBRT; Small fields; Inhomogeneity; Low density

向作者/读者索取更多资源

This review addresses the theme of dose calculation accuracy in the case of the stereotactic treatment of lung lesions. Based on the classical categories of type a, b and c algorithms (according to their degree of complexity in the management of charged particle transport), a summary of findings from literature is reported. Two main critical areas have been identified: the use of small fields and the presence of low density medium. Concerning the latter point, the algorithm accuracy is intrinsic of the algorithm core, and, notwithstanding the materials discretization and their chemical composition knowledge, type c are, at the most, able to reproduce the actual physical dose distribution in heterogeneous media. For what concerns the small field management, the final accuracy could be strongly related to the beam configuration appropriateness in the TPS (as well for MC this relates to the proper linac head description). As a very crude summary, type a should be considered as unsuitable for this kind of treatment calculations (with differences of the order of 20-30%), while type b and c could keep their accuracy approximatively within 10 and 5%, respectively. (C) 2016 Published by Elsevier Ltd on behalf of Associazione Italiana di Fisica Medica.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据