4.6 Article

Benchtop magnetic particle relaxometer for detection, characterization and analysis of magnetic nanoparticles

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 63, 期 17, 页码 -

出版社

IOP PUBLISHING LTD
DOI: 10.1088/1361-6560/aad97d

关键词

magnetic nanoparticle; magnetic particle relaxometer; magnetic particle imaging; relaxation

资金

  1. National Institutes of Health [1R21EB018453-01A1]
  2. University of Florida

向作者/读者索取更多资源

This paper presents the design, construction, and testing of a magnetic particle relaxometer (MPR) to assess magnetic nanoparticle response to dynamic magnetic fields while subjected to a bias field. The designed MPR can characterize magnetic particles for use as tracers in magnetic particle imaging (MPI), with the variation of an applied bias field emulating the scan of the MPI field free point. The system applies a high-frequency time-varying excitation field (up to 45 mT at 30 kHz), while slowly ramping a bias field (+/- 100 mT in 1 s). The time-resolved response of the sample is measured using an inductive sensing coil system, made of a pick-up coil and a rotating and translating balancing coil to finely cancel the induction feed-through from the excitation field. A post-processing algorithm is presented to extract the tracer response related to the point spread function for MPI applications, and the performance of the MPR is demonstrated using superparamagnetic iron oxide particles (ferucarbotran).

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据