3.8 Article

Assessment of Lingual Stability in Mandible Fracture: Monocortical Versus Bicortical Fixation Using FEM Analysis

期刊

JOURNAL OF MAXILLOFACIAL & ORAL SURGERY
卷 17, 期 4, 页码 514-519

出版社

SPRINGER INDIA
DOI: 10.1007/s12663-017-1073-0

关键词

Open reduction and internal fixation (ORIF); Finite element model (FEM) analysis; Monocortical; Bicortical

向作者/读者索取更多资源

Aim of the study The present study compares lingual stability of monocortical miniplate fixation with that of bicortical miniplate fixation in parasymphysis fracture using FEM analysis. Materials and methods Using multislice CT scanner, 3D FEM of patient's mandible was created. Fracture was simulated at parasymphysis region and fixed with 2-mm titanium miniplates and screws of length 8, 10, 12 mm, respectively. Loading force of 120 N applied at molar region and 62.8 N at incisor region. These three models were imported into ANSYS Workbench FEM software. Result There is no significant difference in results between bicortical fixation and monocortical fixation. Conclusion It was concluded that use of monocortical fixation provides sufficient lingual stability. This suggests monocortical fixation system is as reliable as bicortical fixation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据