Reliability of the EchoMRI Infants System for Water and Fat Measurements in Newborns

期刊

OBESITY
卷 25, 期 9, 页码 1577-1583

出版社

WILEY
DOI: 10.1002/oby.21918

关键词

-

资金

  1. National Institutes of Health [P30-DK-26687, UL1 TR000040]
  2. [UO1 DK094463]
  3. [T32-DK007559-25]

向作者/读者索取更多资源

ObjectiveThe precision and accuracy of a quantitative magnetic resonance (EchoMRI Infants) system in newborns were determined. MethodsCanola oil and drinking water phantoms (increments of 10 g to 1.9 kg) were scanned four times. Instrument reproducibility was assessed from three scans (within 10 minutes) in 42 healthy term newborns (12-70 hours post birth). Instrument precision was determined from the coefficient of variation (CV) of repeated scans for total water, lean mass, and fat measures for newborns and the mean difference between weight and measurement for phantoms. In newborns, the system accuracy for total body water (TBW) was tested against deuterium dilution (D2O). ResultsIn phantoms, the repeatability and accuracy of fat and water measurements increased as the weight of oil and water increased. TBW was overestimated in amounts>200g. In newborns weighing 3.14kg, fat, lean mass, and TBW were 0.52kg (16.48%), 2.28kg, and 2.40kg, respectively. EchoMRI's reproducibility (CV) was 3.27%, 1.83%, and 1.34% for total body fat, lean mass, and TBW, respectively. EchoMRI-TBW values did not differ from D2O; mean difference, -1.956.76%, P=0.387; mean bias (limits of agreement), 0.046kg (-0.30 to 0.39kg). ConclusionsThe EchoMRI Infants system's precision and accuracy for total body fat and lean mass are better than established techniques and equivalent to D2O for TBW in phantoms and newborns.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据