Evaluation and benchmarking of an EC-QCL-based mid-infrared spectrometer for monitoring metabolic blood parameters in critical care units

出版社

SPIE-INT SOC OPTICAL ENGINEERING
DOI: 10.1117/12.2289625

关键词

Infrared spectroscopy; quantum cascade lasers; external cavity; blood glucose measurement; multivariate analysis

资金

  1. Ministerium fur Innovation, Wissenschaft und Forschung des Landes NRW
  2. Bundesministerium fur Bildung und Forschung [VIP-03V0433, VIP-03V0434]

向作者/读者索取更多资源

Mid-infrared spectroscopy hyphenated with micro-dialysis is an excellent method for monitoring metabolic blood parameters as it enables the concurrent, reagent-free and precise measurement of multiple clinically relevant substances such as glucose, lactate and urea in micro-dialysates of blood or interstitial fluid. For a marketable implementation, quantum cascade lasers (QCL) seem to represent a favourable technology due to their high degree of miniaturization and potentially low production costs. In this work, an external cavity (EC) - QCL-based spectrometer and two Fourier-transform infrared (FTIR) spectrometers were benchmarked with regard to the precision, accuracy and long-term stability needed for the monitoring of critically ill patients. For the tests, ternary aqueous solutions of glucose, lactate and mannitol (the latter for dialysis recovery determination) were measured in custom-made flow-through transmission cells of different pathlengths and analyzed by Partial Least Squares calibration models. It was revealed, that the wavenumber tuning speed of the QCL had a severe impact on the EC-mirror trajectory due to matching the digital-analog-converter step frequency with the mechanical resonance frequency of the mirror actuation. By selecting an appropriate tuning speed, the mirror oscillations acted as a hardware smoothing filter for the significant intensity variations caused by mode hopping. Besides the tuning speed, the effects of averaging over multiple spectra and software smoothing parameters (Savitzky-Golay-filters and FT-smoothing) were investigated. The final settings led to a performance of the QCL-system, which was comparable with a research FTIR-spectrometer and even surpassed the performance of a small FTIR-mini-spectrometer.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据