4.1 Article

Measurement of phosphate in small samples using capillary electrophoresis with laser-induced luminescence detection

期刊

出版社

TAYLOR & FRANCIS INC
DOI: 10.1080/10826076.2018.1564326

关键词

Capillary electrophoresis; Eu3+-tetracycline; luminescence; phosphate; small sample size; soil

资金

  1. Natural Sciences and Engineering Research Council of Canada

向作者/读者索取更多资源

A capillary electrophoresis-based method was developed for the detection of phosphate in small samples. Phosphate was incubated in buffer with 50 mu M Eu3+-tetracycline to allow for the formation of the luminescent Eu3+-tetracycline-phosphate complex. The complex was separated from the excess Eu3+-tetracycline, which is also luminescent, and the complex detected by post-column laser-induced luminescence within a sheath flow cuvette. Separation time was <90 s with a total run time of 5 min. No flushing of the capillary was required between runs. Phosphate standards ranging from 10-200 mu M were analyzed. Increasing the concentration of Eu3+-tetracycline increased the linear range but at the cost of a decreased sensitivity. A decrease in the Eu3+-tetracycline resulted in an increase in sensitivity down to 5 mu M but a decrease in linear range. As an application, the phosphate content of a milligram sample of soil was determined to be 0.51 ppm. [GRAPHICS] .

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据