4.6 Article

A Miniature Graphene-based Biosensor for Intracellular Glucose Measurements

期刊

ELECTROCHIMICA ACTA
卷 174, 期 -, 页码 574-580

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.electacta.2015.06.035

关键词

Graphene; Bio-sensors; Glucose-oxidase; Intracellular sensors; Graphene oxide; Glucose Sensor

向作者/读者索取更多资源

We report on a small and simple graphene-based potentiometric sensor for the measurement of intracellular glucose concentration. A fine borosilicate glass capillary coated with graphene and subsequently immobilized with glucose oxidase (GOD) enzyme is inserted into the intracellular environment of a single human cell. The functional groups on the edge plane of graphene assist the attachment with the free amine terminals of GOD enzyme, resulting in a better immobilization. The sensor exhibits a glucose-dependent electrochemical potential against an Ag/AgCl reference microelectrode which is linear across the whole concentration range of interest (10 - 1000 mu M). Glucose concentration in human fat cell measured by our graphene-based sensor is in good agreement with nuclear magnetic resonance (NMR) spectroscopy. (C) 2015 Elsevier Ltd. All rights reserved.