4.7 Article

A graphene oxide based poly(2,6-diaminopyridine) composite for solid-phase extraction of Cd(II) prior to its determination by FAAS

期刊

MICROCHIMICA ACTA
卷 184, 期 8, 页码 2855-2860

出版社

SPRINGER WIEN
DOI: 10.1007/s00604-017-2317-x

关键词

Cadmiumin water; Aromatic diamines; In-situ polymerization; FTIR; Preconcentration; Flame atomic absorption spectrometry

向作者/读者索取更多资源

The authors describe the preparation of a graphene oxide based poly(2,6-diaminopyridine) composite by in-situ polymerization. The material was characterized by Fourier transform infrared spectrometry, powder X-ray diffraction and scanning electron microscopy. The composite was placed in a column and used for solid phase extraction (SPE) of Cd(II) ions from water samples. The adsorbed Cd(II) was then eluted with HCl and injected into a flame atomic absorption spectrometer. The effects of pH values, volume and flow rate of sample, and the kind, volume and concentration of eluent were optimized. The calibration plot for the assay is linear in the 2 to 100 mu g.L-1 Cd(II) concentration range. Other figures of merit include (a) a 0.47 mu g.L-1 limit of detection, (b) a relative standard deviation of 1.6% (for n = 7; at 20 mu g.L-1 of Cd(II)), and (c) a preconcentration factor of 50. The method was validated by analyzing the certified reference material NIST SRM 1643f and spiked real samples. The method was successfully employed to the determination of Cd(II) in tap, river, seawater and wastewater samples.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据