4.1 Article

Towards the Implementation of a Wirelessly Powered Dielectric Sensor With Digitized Output for Implantable Applications

期刊

IEEE SENSORS LETTERS
卷 3, 期 3, 页码 -

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/LSENS.2019.2896185

关键词

Sensor systems; chemical and biological sensor; dielectric sensor; energy-harvesting sensor; integrated circuit (IC); wireless power transfer

向作者/读者索取更多资源

This article proposes a wirelessly powered dielectric sensor for industrial and medical applications. The sensor includes a 5-bit binary-weighted capacitor bank and an energy-harvesting rectifier realized as an integrated circuit (IC). With an automated algorithm, the capacitor bank is adjusted to counterbalance the influence of the dielectric material under test. The digitized values of the capacitor bank offer a quantitative measurement of the dielectric constant. The IC was fabricated in 180-nm CMOS process, and the control algorithm was implemented in a discrete microcontroller. The rectifier adopts a differentially cross-connected topology which achieves the rectification efficiency around 72%-82%. The proposed system is used to measure the volume of phosphate-buffered saline (PBS), in which the sensor is placed under the petri dish with PBS samples. This work shows promise for implantable applications such as bladder urine monitoring and other body fluid analysis.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据