3.8 Article

Blood sample quality

期刊

DIAGNOSIS
卷 6, 期 1, 页码 25-31

出版社

WALTER DE GRUYTER GMBH
DOI: 10.1515/dx-2018-0018

关键词

blood samples; errors; hemolysis; laboratory medicine; quality

向作者/读者索取更多资源

Several lines of evidence now confirm that the vast majority of errors in laboratory medicine occur in the extra-analytical phases of the total testing processing, especially in the preanalytical phase. Most importantly, the collection of unsuitable specimens for testing (either due to inappropriate volume or quality) is by far the most frequent source of all laboratory errors, thus calling for urgent strategies for improving blood sample quality and managing data potentially generated measuring unsuitable specimens. A comprehensive overview of scientific literature leads us to conclude that hemolyzed samples are the most frequent cause of specimen nonconformity in clinical laboratories (40-70%), followed by insufficient or inappropriate sample volume (10-20%), biological samples collected in the wrong container (5-15%) and undue clotting (5-10%). Less frequent causes of impaired sample quality include contamination by infusion fluids (i.e. most often saline or glucose solutions), cross-contamination of blood tubes additives, inappropriate sample storage conditions or repeated freezing-thawing cycles. Therefore, this article is aimed to summarize the current evidence about the most frequent types of unsuitable blood samples, along with tentative recommendations on how to prevent or manage these preanalytical non-conformities.