X-ray spectrum estimation for accurate attenuation simulation

期刊

MEDICAL PHYSICS
卷 44, 期 12, 页码 6183-6194

出版社

WILEY
DOI: 10.1002/mp.12607

关键词

computed tomography; detected x-ray spectrum; EM; SVD

资金

  1. AiF [KF2301007NT3]

向作者/读者索取更多资源

Purpose: To estimate detected x-ray spectra from transmission measurements of known attenuators that allow to accurately simulate the transmission in unknown attenuators. Methods: Starting from the established spectrum estimation method using the truncated singular value decomposition (TSVD) we extended the algorithm by incorporating prior knowledge about the statistical nature of the transmission data and about high-frequency spectral components like characteristic peaks. Thereby our proposed approach requires only minimal prior knowledge, namely the energy positions of characteristic peaks or k-edges, which are typically well-known. This ensures that the final spectrum is not biased towards a given prior spectrum which is often observed in other methods. The new approach, prior truncated singular value decomposition (PTSVD), is compared to TSVD as well as the expectation-maximization (EM) method in a simulation and a measurement study. The resulting spectra are evaluated according to their ability to reproduce transmission data of attenuators that have not been included into the estimation process. Results: In case of noiseless simulated data, the PTSVD approach outperforms the existing methods in both, estimating the shape of the spectrum as well as providing a spectrum that reproduces the transmission data. Not surprisingly for increasing noise the ability of PTSVD to estimate the spectral shape worsens while it still performs best in reproducing the transmission data. This finding is also confirmed in the measurement study. Conclusion: Our new approach allows to estimate detected x-ray spectra that accurately reproduce both transmission measurements that have and have not been included into the estimation process. It is less prone to noise compared to the established TSVD method and potentially leads to smaller transmission errors compared to EM for accurate transmission data while being less biased towards the given prior information. (C) 2017 American Association of Physicists in Medicine

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据