4.4 Article

Laser fabrication of electrical feedthroughs in polymer encapsulations for active implantable medical devices

期刊

MEDICAL ENGINEERING & PHYSICS
卷 42, 期 -, 页码 105-110

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.medengphy.2017.01.010

关键词

Hermetic feedthroughs; Bionic implants; Laser joining; Polymer encapsulations; Water vapor transmission rate; Polyetheretherketone

资金

  1. Australian Research Council [LP110200227]
  2. Australian Research Council [LP110200227] Funding Source: Australian Research Council

向作者/读者索取更多资源

Hermetic electrical feedthroughs are essential for safe and functional active implantable biomedical devices and for a wide range of other applications such as batteries, supercapacitors, OLEDs and solar cells. Ceramics and metals have previously been the materials of choice for encapsulations, while polymers have advantages of ease of mass production and end user compatibility. We demonstrate a laser sealing technology that gives hermetic, mechanically strong feedthroughs with low electrical resistance in a polyetheretherketone (PEEK) encapsulation. The conductive pathways are wires and sputtered thin films. The water vapor transmission rate through the fabricated encapsulations is comparable to that of PEEK itself. (C) 2017 Published by Elsevier Ltd on behalf of IPEM.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。