4.5 Article

High-sensitivity operation of single-beam optically pumped magnetometer in a kHz frequency range

期刊

MEASUREMENT SCIENCE AND TECHNOLOGY
卷 28, 期 3, 页码 -

出版社

IOP PUBLISHING LTD
DOI: 10.1088/1361-6501/aa58b4

关键词

atomic magnetometer; fT sensitivity; optically pumped magnetometer; single-beam configuration; high-frequency operation

资金

  1. US DOE through the LANL/LDRD program

向作者/读者索取更多资源

Optically pumped magnetometers (OPM) can be used in various applications, from magnetoencephalography to magnetic resonance imaging and nuclear quadrupole resonance (NQR). OPMs provide high sensitivity and have the significant advantage of non-cryogenic operation. To date, many magnetometers have been demonstrated with sensitivity close to 1 fT, but most devices are not commercialized. Most recently, QuSpin developed a model of OPM that is low cost, high sensitivity, and convenient for users, which operates in a singlebeam configuration. Here we developed a theory of single-beam (or parallel two-beam) magnetometers and showed that it is possible to achieve good sensitivity beyond their usual frequency range by tuning the magnetic field. Experimentally we have tested and optimized a QuSpin OPM for operation in the frequency range from DC to 1.7 kHz, and found that the performance was only slightly inferior despite the expected decrease due to deviation from the spin-exchange relaxation-free regime.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据