A new finding on the in-vivo crevice corrosion damage in a CoCrMo hip implant

出版社

ELSEVIER
DOI: 10.1016/j.msec.2017.05.086

关键词

Crevice corrosion; CoCrMo alloy; Chromium oxide; Fretting wear; Orthopaedic implants; X-ray photoelectron spectroscopy

资金

  1. Premier's Research and Industry Fund's Catalyst Research Grant, Department of State Development, Government of South Australia [CRG 46]

向作者/读者索取更多资源

A detailed investigation was performed to characterize the fretting wear and corrosion damage to the neck component of a CoCrMo stem from a metal-on-polyethylene implant retrieved after 99 months. The stem was a low carbon (0.07 wt%) wrought Co-28Cr-6Mo alloy with no secondary carbide phases in the matrix (gamma-phase). The original design of the neck surface contained an intentionally fabricated knurled profile with a valley-to-peak range of approximately 11 mu m. Roughness measurements indicated that the tip of the knurled profile was significantly damaged, especially in the distal medial region of the neck, with up to a 22% reduction in the mean peak to -valley height (R-a) compared to the original profile. As a new finding, the channels between the peaks of the profile created an additional crevice site in the presence of stagnant body fluid within the head-neck taper junction. These channels were observed to contain the most severe corroded areas and surface oxide layers with micro-cracks. SEM/EDS, XRD and XPS evaluations identified the formation of Cr2O3 as a corrosion product. Also, decobaltification was found to occur in these corroded areas. The findings of this work indicate the important role of the knurled profile in inducing additional crevice corrosion. (C) 2017 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据