3D phase contrast MRI: Partial volume correction for robust blood flow quantification in small intracranial vessels

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 79, 期 1, 页码 129-140

出版社

WILEY
DOI: 10.1002/mrm.26637

关键词

Partial volume correction; blood flow; PCMRI; intracranial arteries

资金

  1. Swiss National Science Foundation [SNF 32003B_160222, SNF 320030_156813]
  2. Swiss National Science Foundation (SNF) [32003B_160222, 320030_156813] Funding Source: Swiss National Science Foundation (SNF)

向作者/读者索取更多资源

PurposeRecent advances in 3D-PCMRI (phase contrast MRI) sequences allow for measuring the complex hemodynamics in cerebral arteries. However, the small size of these vessels vs spatial resolution can lead to non-negligible partial volume artifacts, which must be taken into account when computing blood flow rates. For this purpose, we combined the velocity information provided by 3D-PCMRI with vessel geometry measured with 3DTOF (time of flight MRI) or 3DRA (3D rotational angiography) to correct the partial volume effects in flow rate assessments. MethodsThe proposed methodology was first tested in vitro on cylindrical and patient specific vessels subject to fully controlled pulsatile flows. Both 2D- and 3D-PCMRI measurements using various spatial resolutions ranging from 20 to 1.3 voxels per vessel diameter were analyzed and compared with flowmeter baseline. Second, 3DTOF, 2D- and 3D-PCMRI measurements were performed in vivo on 35 patients harboring internal carotid artery (ICA) aneurysms indicated for endovascular treatments requiring 3DRA imaging. ResultsThe in vitro 2D- and 3D-PCMRI mean flow rates assessed with partial volume correction showed very low sensitivity to the acquisition resolution above approximate to 2 voxels per vessel diameter while uncorrected flow rates deviated critically when decreasing the spatial resolution. 3D-PCMRI flow rates measured in vivo in ICA agreed very well with 2D-PCMRI data and a good flow conservation was observed at the C7 bifurcation. Globally, partial volume correction led to 10-15% lower flow rates than uncorrected values as those reported in most of the published studies on intracranial flows. ConclusionPartial volume correction may improve the accuracy of PCMRI flow rate measurements especially in small vessels such as intracranial arteries. Magn Reson Med 79:129-140, 2018. (c) 2017 International Society for Magnetic Resonance in Medicine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据