4.5 Article

Quantitative magnetization transfer ultrashort echo time imaging using a time-efficient 3D multispoke Cones sequence

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 79, 期 2, 页码 692-700

出版社

WILEY
DOI: 10.1002/mrm.26716

关键词

ultrashort echo time; cones; magnetization transfer; two-pool modeling; short T-2 tissues

资金

  1. GE Healthcare
  2. NIH [1R01 AR062581, 1R01 AR068987]
  3. VA Clinical Science RD Service [1I01CX001388]

向作者/读者索取更多资源

PurposeTo accelerate the quantitative ultrashort echo time imaging using a time-efficient 3D multispoke Cones sequence with magnetization transfer (3D UTE-Cones-MT) and signal modeling. Theory and MethodsA 3D UTE-Cones-MT acquisition scheme with multispoke per MT preparation and a modified rectangular pulse (RP) approximation was developed for two-pool MT modeling of macromolecular and water components including their relative fractions, relaxation times and exchange rates. Numerical simulation and cadaveric specimens, including human Achilles tendon and bovine cortical bone, were investigated using a clinical 3T scanner. ResultsNumerical simulation showed that the modified RP model provided accurate estimation of MT parameters when multispokes were acquired per MT preparation. For the experiment with the Achilles tendon and cortical bone samples, the macromolecular fractions were 20.42.0% and 59.4 +/- 5.3%, respectively. ConclusionThe 3D multispoke UTE-Cones-MT sequence can be used for fast volumetric assessment of macromolecular and water components in short T-2 tissues. Magn Reson Med 79:692-700, 2018. (c) 2017 International Society for Magnetic Resonance in Medicine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据