4.5 Article

Validation of highly accelerated real-time cardiac cine MRI with radial k-space sampling and compressed sensing in patients at 1.5T and 3T

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 79, 期 5, 页码 2745-2751

出版社

WILEY
DOI: 10.1002/mrm.26918

关键词

compressed sensing; radial k-space; Cartesian k-space; real-time cine MRI; tiny golden angles; cardiac function

资金

  1. National Institutes of Health [R01HL116895, R01HL138578, R21EB024315, R21AG055954, R01DK102438]

向作者/读者索取更多资源

PurposeTo validate an optimal 12-fold accelerated real-time cine MRI pulse sequence with radial k-space sampling and compressed sensing (CS) in patients at 1.5T and 3T. MethodsWe used two strategies to reduce image artifacts arising from gradient delays and eddy currents in radial k-space sampling with balanced steady-state free precession readout. We validated this pulse sequence against a standard breath-hold cine sequence in two patient cohorts: a myocardial infarction (n=16) group at 1.5T and chronic kidney disease group (n=18) at 3T. Two readers independently performed visual analysis of 68 cine sets in four categories (myocardial definition, temporal fidelity, artifact, noise) on a 5-point Likert scale (1=nondiagnostic, 2=poor, 3=adequate or moderate, 4=good, 5=excellent). Another reader calculated left ventricular (LV) functional parameters, including ejection fraction. ResultsCompared with standard cine, real-time cine produced nonsignificantly different visually assessed scores, except for the following categories: 1) temporal fidelity scores were significantly lower (P=0.013) for real-time cine at both field strengths, 2) artifacts scores were significantly higher (P=0.013) for real-time cine at both field strengths, and 3) noise scores were significantly (P=0.013) higher for real-time cine at 1.5T. Standard and real-time cine pulse sequences produced LV functional parameters that were in good agreement (e.g., absolute mean difference in ejection fraction <4%). ConclusionThis study demonstrates that an optimal 12-fold, accelerated, real-time cine MRI pulse sequence using radial k-space sampling and CS produces good to excellent visual scores and relatively accurate LV functional parameters in patients at 1.5T and 3T. Magn Reson Med 79:2745-2751, 2018. (c) 2017 International Society for Magnetic Resonance in Medicine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据