4.5 Article

Respiratory motion model based on the noise covariance matrix of a receive array

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 79, 期 3, 页码 1730-1735

出版社

WILEY
DOI: 10.1002/mrm.26775

关键词

respiratory motion model; noise covariance matrix; noise sensor; tracking

资金

  1. ITEA [12026]
  2. European Research Council [ERC-2010-AdG-20100317]

向作者/读者索取更多资源

PurposeTracking of the internal anatomy by means of a motion model that uses the MR-derived motion fields and noise covariance matrix (NCM) dynamic as a surrogate signal. MethodsA 2D respiratory motion model was developed based on the MR-derived motion fields and the NCM of a receive array used in MRI. Temporal dynamics of the NCM were used as a motion surrogate for a linear correspondence motion model. The model performance was tested on five healthy volunteers with a liver as the target. The motion fields were calculated from the cineMR frames with an optical flow registration tool. ResultsThe model estimated the liver motion with an average residual error of 2.3mm (13% of the motion amplitude). The model formation takes 3min and the model latency was 0.5 s in the current implementation. The limiting factor for the latency is the current update time of the NCM (0.48 s), which in principle can be reduced to 0.004 s with an alternative way to determine the NCM. ConclusionsThe 2D respiratory motion of the liver can be effectively estimated with the linear motion model that uses the temporal behavior of the NCM as motion surrogate. Magn Reson Med 79:1730-1735, 2018. (c) 2017 International Society for Magnetic Resonance in Medicine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据