4.3 Article

Coagulated region analysis in a microwave surgical device with temperature-dependent physical properties of tissue

期刊

IEICE ELECTRONICS EXPRESS
卷 16, 期 9, 页码 -

出版社

IEICE-INST ELECTRONICS INFORMATION COMMUNICATIONS ENG
DOI: 10.1587/elex.16.20190131

关键词

microwave surgical device; coagulated region; physical properties of tissue

向作者/读者索取更多资源

In surgical devices, it is pointed out that tissues around these devices may experience thermal damage ( coagulation). Therefore, evaluating the coagulated region generated by a microwave surgical device is an important issue. In this paper, we estimate the distribution of the coagulated region using numerical calculation. Then, there is a possibility that the calculated results will be affected by change to the electrical and thermal constants during heating. Therefore, the temperature dependence of the tissue was considered. As a result, it was confirmed that the coagulation width at the center of the lower blade was 0.5 mm larger than that without temperature dependence.