Calibration strategies to correct for matrix effects in direct analysis of urine by ICP OES: internal standardization and multi-energy calibration

期刊

ANALYTICAL METHODS
卷 11, 期 27, 页码 3401-3409

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c9ay00907h

关键词

-

资金

  1. Conselho Nacional de Desenvolvimento Cientico e Tecnologico (CNPq) [141634/2017-0, 303107/2013-8, 305201/2018-2, 428558/2018-6]
  2. Coordenacao de Aperfeicoamento de Pessoal de Nivel Superior (CAPES/PNPD - Graduate Program in Chemistry, Federal University of Sao Carlos)
  3. Coordenacao de Aperfeicoamento de Pessoal de Nivel Superior - Brasil (CAPES) [001]
  4. Analitica (Sao Paulo, SP, Brazil)
  5. Thermo Scientific

向作者/读者索取更多资源

This study investigated internal standardization and multi-energy calibration (MEC) as strategies for direct analysis of urine by inductively coupled plasma optical emission spectrometry (ICP OES). Gallium, Ge, Pd, Sc and Y were evaluated as internal standards (ISs) for determination of Al, As, Ba, Be, Bi, Ca, Cd, Co, Cr, Cu, Hg, K, Li, Mg, Na, Pb, Sb and Zn in urine. The accuracy and precision were evaluated by addition-recovery experiments and the best results were obtained when Ge or Pd was used as IS. Recoveries ranged from 80 to 120% and relative standard deviations for all elements and samples were lower than 7.5%, except in the case of the lowest analyte levels (0.025 and 0.050 mg L-1). Recoveries varying from 80 to 120% were obtained only for the highest levels tested (>4.0 mg L-1) when using MEC. The LODs obtained by internal standardization were lower than the LODs obtained by MEC. Both calibration methods were suitable to correct for matrix effects, making feasible the direct analysis of urine by ICP OES, without any sample preparation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据