期刊
BIOINTERFACE RESEARCH IN APPLIED CHEMISTRY
卷 9, 期 4, 页码 4022-4026出版社
BIOINTERFACE RESEARCH APPLIED CHEMISTRY
DOI: 10.33263/BRIAC94.022026
关键词
Biosensor; Nanofiber; Electrospinning; Carbon
资金
- Tehran University of Medical Sciences (TUMS) [96-01-87-34100]
The purpose of this work was to design the electrochemical DNA biosensor based on carbon nanofibers (CNFs) for detecting hepatitis B virus (HBV). The CNFs, due to high conductivity and surface-to-volume ratio, was known as an effective material in the electrochemical biosensor. In this work, we directly used electrospun CNF as the electrode. CNF electrode was modified with electropolymerized glutamic acid (Glu). Then, the probe DNA (pDNA) was conjugated to Glu modified CNFs electrode. The surface morphology of CNFs and Glu modified CNFs were characterized via scanning electron microscopy (SEM). Electropolymerized Glu on the surface of CNF electrode was characterized by Fourier transform infrared spectroscopy (FTIR) as well. The cyclic voltammetry (CV) was used to monitor the target DNA (tDNA). The tDNA was quantified at a linear range from 1 x 10(-12) to 1 x 10(-6) M with a detection limit of 1.58 x 10(-12) M. This electrochemical HBV biosensor had good stability, repeatability, and selectivity for distinguishing complementary DNA from non-complementary and mis-matched DNA sequence.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐