3.8 Proceedings Paper

Magnetic graphene oxide for efficient solid phase extraction of DEHP

出版社

IOP PUBLISHING LTD
DOI: 10.1088/1757-899X/544/1/012063

关键词

-

向作者/读者索取更多资源

DEHP (di-2-ethylhexyl phthalate) can cause negative impacts on human health, thus development of reliable analytical methods for the determination of PAEs is important. In this work, magnetic graphene oxide was synthesized and used as the sorbent for solid phase extraction of DEHP. Factors that influence the extraction efficiency of DEHP were studied. The optimum conditions, such as solution pH, extraction time, desorption time, and eluent type and volume were obtained. The analytical performance of the developed method was investigated, and it was found that the limit of detection (LOD) for DEHP was 0.35 mu g/L and the related standard deviation (RSD) was 4.3 %. The proposed method was applied in the determination of DEHP in environmental water samples with recoveries of 91.6-106.5%.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据